FDA Device Recall Z-3086-2017
Maquet Datascope Corp - Cardiac Assist Division · Mahwah, NJ
Class I — life-threatening Ongoing
Device
CS 100 Intra-Aortic Balloon Pump. It supports the heart's left ventricle by increasing coronary perfusion and reducing left ventricular work. Coronary perfusion is increased by augmenting blood pressure during the diastolic phase of the cardiac cycle. This increase in aortic pressure promotes more blood flow through the coronary arteries. Left ventricular work is reduced by decreasing aortic end diastolic pressure and reducing resistance to ventricular ejection, resulting in a decrease in blood pressure during the systolic phase of the cardiac cycle.
Reason for recall
False blood back detection alarm and the ingress of fluids into the IABP affecting various electronic circuit boards could potentially prevent initiation or continuation of therapy.
Recall record
- Recall number
Z-3086-2017- Classification
- Class I
- Status
- Ongoing
- Voluntary or mandated
- Voluntary: Firm initiated
- Firm notification
- Letter
- Distribution
- Worldwide Distribution - US Nationwide International: AFGHANISTAN, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BANGLADESH, BELARUS, BELGIUM, BOLIVIA, BOSNIA-HERZEGOVINA, BRAZIL, BRUNEI, BULGARIA, CAMBODIA, CANADA, CHILE, CHINA, COLOMBIA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, GUATEMALA, GUYANA, HONDURAS, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KOREA, KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MALDIVES, MALTA, MAURITIUS, MEXICO, MOROCCO, MYANMAR, NEPAL, NETHERLANDS, NEW ZEALAND, NORTH KOREA, NORWAY, OMAN, PAKISTAN, PALESTINE, PANAMA, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, TANZANIA, THAILAND, TRINIDAD, TUNISIA, TURKEY, TURKMENISTAN, UGANDA, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, UZBEKISTAN, VENEZUELA, VIETNAM & YEMEN.
- Recall initiated
- 2017-07-17
- Classified by FDA Center
- 2017-09-05
- FDA published
- 2017-09-13
- Recalling firm
- Maquet Datascope Corp - Cardiac Assist Division
- Firm location
- Mahwah, NJ
Operational response
A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.