FDA Device Recall Z-2690-2024
Breas Medical, Inc. · North Billerica, MA
Class I — life-threatening Ongoing
Device
Vivo 45 LS, pressure and volume ventilator capable of delivering continuous or intermittent ventilatory support for patients who require invasive or non-invasive mechanical ventilation, Model/Catalog Number: 230000 (230016, 230216)
Reason for recall
There is a potential for short term (<7 days) elevated levels of formaldehyde emitted from the device into the patient breathing airpath in specific conditions.
Recall record
- Recall number
Z-2690-2024- Classification
- Class I
- Status
- Ongoing
- Voluntary or mandated
- Voluntary: Firm initiated
- Firm notification
- Letter
- Distribution
- US Nationwide.
- Recall initiated
- 2024-07-25
- Classified by FDA Center
- 2024-09-03
- FDA published
- 2024-09-11
- Recalling firm
- Breas Medical, Inc.
- Firm location
- North Billerica, MA
Operational response
A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.
For the official FDA enforcement record, see FDA's Recall Search.