FDA Device Recall Z-2661-2025

Siemens Medical Solutions USA, Inc · Malvern, PA

Class I — life-threatening Ongoing

Highest impact — Class I device recall — FDA determination that use of, or exposure to, the device will cause serious adverse health consequences or death. Distribution is nationwide and the recall remains in progress.

Device

MAGNETOM Skyra fit. Model Number: 10849580.

Lot / serial / GTIN: Model Number: 10849580. UDI Numbers: (01)04056869006710(21)170068, (01)04056869006710(21)170053, (01)04056869006710(21)70022, (01)04056869006710(21)170049, (01)04056869006710(21)170018, (01)04056869006680(21)162078, (01)04056869006710(21)170109, (01)04056869006710(21)170039, (01)04056869006710(21)170066, (01)04056869006710(21)170023, (01)04056869006710(21)170030, (01)04056869006710(21)170004, (01)04056869006710(21)170016, (01)04056869006710(21)170077, (01)04056869006710(21)170058, (01)04056869006710(21)170002, (01)04056869006710(21)170050, (01)04056869006710(21)170010, (01)04056869006710(21)170019, (01)04056869006710(21)70026, (01)04056869006710(21)170056, (01)04056869006710(21)170061, (01)04056869006710(21)170105, (01)04056869006710(21)170017, (01)04056869006710(21)170041, (01)04056869006710(21)70032, (01)04056869006710(21)170111, (01)04056869006710(21)170048, (01)04056869006710(21)170101, (01)04056869006710(21)170100, (01)04056869006710(21)170060, (01)04056869006710(21)170099, (01)04056869006710(21)170036, (01)04056869006710(21)170007, (01)04056869006710(21)170001, (01)04056869006710(21)170009, (01)04056869006710(21)70020, (01)04056869006710(21)170112, (01)04056869006710(21)170113, (01)04056869006710(21)170054, (01)04056869006710(21)170047, (01)04056869006710(21)170046, (01)04056869006710(21)70023, (01)04056869006710(21)70027, (01)04056869006710(21)170110, (01)04056869006710(21)170095, (01)04056869006710(21)170075, (01)04056869006710(21)170011, (01)04056869006710(21)170081, (01)04056869006710(21)170044, (01)04056869006710(21)70028, (01)04056869006710(21)170043, (01)04056869006710(21)70031, (01)04056869006710(21)70008. Serial Numbers: 170068, 170053, 70022, 170049, 170018, 170106, 170109, 170039, 170066, 170023, 170030, 170004, 170016, 170077, 170058, 170002, 170050, 170010, 170019, 70026, 170056, 170061, 170105, 170017, 170041, 70032, 170111, 170048, 170101, 170100, 170060, 170099, 170036, 170007, 170001, 170009, 70020, 170112, 170113, 170054, 170047, 170046, 70023, 70027, 170110, 170095, 170075, 170011, 170081, 170044, 70028, 170043, 70031, 70008. Additional serial number added 2/25/2026: 170023. Additional serial Numbers Added 04/16/2026: 170013.

Quantity: 56 units

Reason for recall

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

Recall record

Recall number
Z-2661-2025
Classification
Class I
Status
Ongoing
Voluntary or mandated
Voluntary: Firm initiated
Firm notification
Letter
Distribution
Worldwide distribution - US Nationwide and the countries of United Arab Emirates, Armenia, Angola, Argentina, Austria, Australia, Azerbaijan, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bahrain, Brunei Darussalam, Brazil, Botswana, Belarus, Canada, Switzerland, Chile, China, Colombia, Cyprus, Czechia, Germany, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Ethiopia, Finland, Faroe Islands, France, United Kingdom of Great Britian and Northern Ireland, Georgia, Greece, Hong Kong, Croatia, Hungary, Canary Islands, Indonesia, Ireland, Israel, India, Iran, Iceland, Italy, Jordan, Japan, Kenya, Korea, Kuwait, Kazakhstan, Lebanon, Sri Lanka, Lithuania, Luxembourg, Latvia, Morocco, Monaco, Moldova, Montenegro, Republic of North Macedonia, Myanmar, Mongolia, Macao, Malta, Maldives, Mexico, Malaysia, Mozambique, Nigeria, Netherlands, Norway, Nepal, New Zealand, Oman, Panama, Peru, Philippines, Pakistan, Poland, Portugal, Paraguay, Qatar, Reunion, Romania, Serbia, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Senegal, Syrian Arab Republic, Thailand, Turkmenistan, Tunisia, Turkey, Trinidad and Tobago, Taiwan, Tanzania, Ukraine, Uganda, Uzbekistan, Viet Nam, Kosovo, South Africa.
Recall initiated
2025-08-28
Classified by FDA Center
2025-09-30
FDA published
2025-10-08
Recalling firm
Siemens Medical Solutions USA, Inc
Firm location
Malvern, PA

Operational response

A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.

For the official FDA enforcement record, see FDA's Recall Search.