FDA Device Recall Z-2610-2025
Medline Industries, LP · Northfield, IL
Class I — life-threatening Ongoing
Device
Reprocessed Livewire Steerable Electrophysiology Catheter Item Numbers 401575RH 401582RH 401600RH 401603RH 401606RH 401652RH 401653RH 401654RH 401904RH 401905RH 401908RH 401914RH 401915RH 401918RH 401932RH 401933RH 401934RH 401938RH 401939RH 401940RH 401941RH 401991RH
Reason for recall
Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
Recall record
- Recall number
Z-2610-2025- Classification
- Class I
- Status
- Ongoing
- Voluntary or mandated
- Voluntary: Firm initiated
- Firm notification
- Letter
- Distribution
- Domestic US distribution nationwide. International distribution pending. No international distribution.
- Recall initiated
- 2025-07-31
- Classified by FDA Center
- 2025-09-26
- FDA published
- 2025-10-08
- Recalling firm
- Medline Industries, LP
- Firm location
- Northfield, IL
Operational response
A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.
For the official FDA enforcement record, see FDA's Recall Search.