FDA Device Recall Z-2368-2023

Covidien LP · Mansfield, MA

Class I — life-threatening Ongoing

Highest impact — Class I device recall — FDA determination that use of, or exposure to, the device will cause serious adverse health consequences or death. Distribution is nationwide and the recall remains in progress.

Device

MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 16 cm, Curved Extensions, IC Tray REF: 8888101002HP

Lot / serial / GTIN: GTIN: 20884521127965 LOT Numbers: 2133600171 2133600172 2133700047 2133700048 2133700060 2200400110 2204000236 2230400213 2230400214 2233400068

Quantity: 6615

Reason for recall

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Recall record

Recall number
Z-2368-2023
Classification
Class I
Status
Ongoing
Voluntary or mandated
Voluntary: Firm initiated
Firm notification
Letter
Distribution
US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom
Recall initiated
2023-06-28
Classified by FDA Center
2023-08-11
FDA published
2023-08-23
Recalling firm
Covidien LP
Firm location
Mansfield, MA

Operational response

A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.

For the official FDA enforcement record, see FDA's Recall Search.