FDA Device Recall Z-2028-2023
GE Healthcare Finland Oy · Helsinki, N/A
Class I — life-threatening Ongoing
Device
TruSignal Sensitive Skin Sensor, REF TS-SE-3; Oximeter
Reason for recall
There is a potential reduction of energy reaching patient during defibrillation, potential contact with unintended voltage, or inaccurate measurement.
Recall record
- Recall number
Z-2028-2023- Classification
- Class I
- Status
- Ongoing
- Voluntary or mandated
- Voluntary: Firm initiated
- Firm notification
- Letter
- Distribution
- Worldwide
- Recall initiated
- 2023-05-19
- Classified by FDA Center
- 2023-07-06
- FDA published
- 2023-07-12
- Recalling firm
- GE Healthcare Finland Oy
- Firm location
- Helsinki, N/A, Finland
Operational response
A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.
For the official FDA enforcement record, see FDA's Recall Search.