Tandem Diabetes Care, Inc. · San Diego, CA

Class I — life-threatening Ongoing

Highest impact — Class I device recall — FDA determination that use of, or exposure to, the device will cause serious adverse health consequences or death.

Device

t:slim X2 Insulin Pump with Interoperable Technology

Lot / serial / GTIN: Software versions 7.9/UDI: 00389152407012

Quantity: 63 units

Reason for recall

A software defect in Version 7.9 of the pump software for Tandem t:slim X2 and Tandem Mobi pumps, when used with Control IQ+ technology, will cause the pump to incorrectly interpolate glucose trends when the Estimated Glucose Value (EGV) is above 255 mg/dL at the start or end of a gap in data collection due to a lapse in connection from a paired continuous glucose monitor (CGM) sensor, which can lead to under-delivery or over-delivery of insulin based on inaccurate result leading to severe cases of hypoglycemia or hyperglycemia.

Recall record

Recall number
Z-1519-2025
Classification
Class I
Status
Ongoing
Voluntary or mandated
Voluntary: Firm initiated
Firm notification
Telephone
Distribution
US: AZ, CA, CO, CT, FL, HI, IA, ID, IL, IN, MD, ME, MI, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, PA, TX, UT, VA, WA, WI OUS: N/A
Recall initiated
2025-02-27
Classified by FDA Center
2025-04-11
FDA published
2025-04-23
Recalling firm
Tandem Diabetes Care, Inc.
Firm location
San Diego, CA

Operational response

A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.

For the official FDA enforcement record, see FDA's Recall Search.