BRANMOOR
THURSDAY · 14 MAY 2026

FDA Device Recall Z-1074-2025

Getinge Usa Sales Inc · Wayne, NJ

Class I — life-threatening Ongoing

Highest impact — Class I device recall — FDA determination that use of, or exposure to, the device will cause serious adverse health consequences or death.

Device

Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i C20, Flow-i C30, Flow-i C40, Flow-c, Flow-e. For use during anesthesia.

Lot / serial / GTIN: Model Number: 6682285. UDI-DI: 07325710001141. Serial Numbers: 3761, 7635, 7633, 7630, 7632, 7631.

Quantity: 6 units

Reason for recall

Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.

Recall record

Recall number
Z-1074-2025
Classification
Class I
Status
Ongoing
Voluntary or mandated
Voluntary: Firm initiated
Firm notification
Letter
Distribution
Domestic distribution to Kentucky and Connecticut.
Recall initiated
2025-01-15
Classified by FDA Center
2025-02-07
FDA published
2025-02-19
Recalling firm
Getinge Usa Sales Inc
Firm location
Wayne, NJ

Operational response

A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.

For the official FDA enforcement record, see FDA's Recall Search.

‹ All device recalls