FDA Device Recall Z-1074-2025
Getinge Usa Sales Inc · Wayne, NJ
Class I — life-threatening Ongoing
Device
Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i C20, Flow-i C30, Flow-i C40, Flow-c, Flow-e. For use during anesthesia.
Reason for recall
Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.
Recall record
- Recall number
Z-1074-2025- Classification
- Class I
- Status
- Ongoing
- Voluntary or mandated
- Voluntary: Firm initiated
- Firm notification
- Letter
- Distribution
- Domestic distribution to Kentucky and Connecticut.
- Recall initiated
- 2025-01-15
- Classified by FDA Center
- 2025-02-07
- FDA published
- 2025-02-19
- Recalling firm
- Getinge Usa Sales Inc
- Firm location
- Wayne, NJ
Operational response
A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.
For the official FDA enforcement record, see FDA's Recall Search.