FDA Device Recall Z-1070-2025

Spectranetics Corporation · Colorado Springs, CO

Class I — life-threatening Ongoing

Highest impact — Class I device recall — FDA determination that use of, or exposure to, the device will cause serious adverse health consequences or death. Distribution is nationwide and the recall remains in progress.

Device

Intact Vascular Tack Endovascular System, to treat vascular dissections with Tack implant(s) following angioplasty: (4F,1.5-4.5mm), 150cm, REF: 154150041; Tack, 4F Gen 1.5, 150cm CE, REF: 154150042; Tack, 4F Gen 1.5, 90cm CE, REF: 154090042 (6F, 3.5 - 6.0mm), 135cm, REF: 156135061; (6F, 4.0 - 8.0mm), 135cm, REF: 206135061; Tack, 6F Gen 2.0, 135cm CE, REF: 206135062; Tack, 6F Gen 1.5, 135cm CE, REF: 156135062; Tack, 6F Gen 2.0, 80cm CE, REF: 206080062; Tack, 6F Gen 1.5, 80cm CE, REF: 156080062

Lot / serial / GTIN: UDI-DI: REF/UDI-DI/Batch: 154150041/00863328000189/332905 - 349049, 154150042/328641, REF: 154090042/322101, 156135061/00863328000103/332824 - 336121, 206135061/00850003494043/331281 - 336152, 206135062/328643 - 345567, 156135062/271330 - 283066, 206080062/321083 - 349035, 156080062/302251 - 349034

Quantity: 2,939

Reason for recall

Use of Tack Endovascular system, designed to treat acute dissections of inner wall or lining of an artery by tacking damaged tissue to inner luminal surface, could result in problems with deployment and stability of device in vessel after deployment, which may cause failure to resolve dissection, migration of implant, bailout stenting, reintervention, unintended removal of tack devices, and ischemia.

Recall record

Recall number
Z-1070-2025
Classification
Class I
Status
Ongoing
Voluntary or mandated
Voluntary: Firm initiated
Firm notification
N/A
Distribution
Distribution US nationwide and countries of: NL, DE, NL, AU, IT, CH, GB, and AT
Recall initiated
2025-01-10
Classified by FDA Center
2025-02-14
FDA published
2025-02-26
Recalling firm
Spectranetics Corporation
Firm location
Colorado Springs, CO

Operational response

A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.