FDA Device Recall Z-0957-2024

Wipro GE Healthcare Private Ltd. · Bengaluru, N/A

Class I — life-threatening Ongoing

Highest impact — Class I device recall — FDA determination that use of, or exposure to, the device will cause serious adverse health consequences or death.

Device

GE Healthcare neonatal incubator, models: a) Care Plus, b) Care Plus 1000, c) Care Plus 2000, d) Care Plus 3000, e) Care Plus 4000, Plus 4000

Lot / serial / GTIN: a) Care Plus, Model Numbers: 6600-0040-900, 6600-0203-900, 6600-0203-903, 6600-0232-910, 6600-0232-920, 6600-0235-901, CARE-PLUS; b) Care Plus 1000, Model Numbers: 8200, 6600-0264-901, 6600-0264-903, 6600-0265-901, 6600-0265-902, 6600-0265-903, 6600-0265-908, 6600-0266-901, 6600-0266-903, 6600-0267-901, 6600-0269-901, 6600-0269-903, 6600-0270-901, 6600-0272-901; c) Care Plus 2000, Model Numbers: 2073175-001, 6600-0273-901, 6600-0274-901, 6600-0274-903, 6600-0274-911, 6600-0275-901, 6600-0275-902, 6600-0275-903, 6600-0275-908, 6600-0275-911, 6600-0276-901, 6600-0276-903, 6600-0277-901, 6600-0279-903, 6600-0280-901, 6600-0280-903, 6600-0281-903, 6600-0282-901, M1050218, M1160968, M1206249, M1242714, QCR043; d) Care Plus 3000, Model Numbers: 6600-0203-902, 6600-0223-981, 6600-0230-900, 6600-0230-901, 6600-0232-901, 6600-0232-911, 6600-0232-921, 6600-0232-980, 6600-0232-983, 6600-0233-900, 6600-0233-901, 6600-0233-911, 6600-0235-900, 6600-0293-901, 6600-0293-903, 6600-0294-901, 6600-0294-902, 6600-0294-903, 6600-0294-905, 6600-0294-908, 6600-0294-909, 6600-0295-901, 6600-0301-901, 6600-0301-903, 6600-0302-901, 6600-0302-903, 6600-0303-901, 6600-0333-903, 6600-0463-901, 6600-0511-901, 6600-0514-900, 6623-0010-903, 6623-0010-923, 6623-0011-902, 6623-0011-903, 6623-0011-920, 6623-0011-923, 6623-0021-920, 6623-0111-900, 6623-0111-903, 6623-0111-905, 6623-0311-900, 6623-0311-903, 6623-0411-900, 6623-0411-905, 6623-0511-900, 6623-0511-902, 6623-0511-903, 6623-0511-905, 6623-0511-910, 6623-0511-913, 6623-0511-920, 6623-0511-921, 6623-0511-943, 6623-0521-900, 6623-0521-902, 6623-0521-905, 6623-0521-910, 6623-0521-943, 6623-0610-905, 6623-0611-900, 6623-0611-903, 6623-0611-905, 6623-0611-943, 6623-0621-943, 6623-0711-900, 6623-0711-903, 6623-0711-920, 6623-0811-900, 6623-0811-903, 6623-0811-910, 6623-0811-913, 6623-0811-920, 6623-0811-943, 6623-0911-900, 6623-0911-905, 6623-0921-905, 6623-1011-903, 6623-1411-900, 6623-1421-905, 6623-1511-900, 6626-0010-903, 6626-0010-913, 6626-0010-923, 6626-0011-900, 6626-0011-902, 6626-0011-903, 6626-0011-905, 6626-0011-920, 6626-0011-923, 6626-0021-903, 6626-0111-900, 6626-0121-905, 6626-0121-920, 6626-0211-900, 6626-0221-900, 6626-0221-903, 6626-0311-900, 6626-0311-902, 6626-0311-903, 6626-0311-905, 6626-0311-910, 6626-0321-900, 6626-0411-900, 6626-0411-903, 6626-0411-905, 6626-0411-920, 6626-0421-900, 6626-0421-913, 6626-0421-920, 6626-0510-900, 6626-0510-905, 6626-0511-900, 6626-0511-901, 6626-0511-902, 6626-0511-903, 6626-0511-905, 6626-0511-910, 6626-0511-913, 6626-0511-920, 6626-0511-922, 6626-0511-923, 6626-0521-900, 6626-0521-902, 6626-0521-903, 6626-0521-905, 6626-0521-913, 6626-0521-920, 6626-0521-923, 6626-0610-923, 6626-0611-900, 6626-0611-903, 6626-0611-905, 6626-0611-920, 6626-0611-923, 6626-0621-903, 6626-0710-903, 6626-0711-900, 6626-0711-903, 6626-0711-905, 6626-0810-923, 6626-0811-900, 6626-0811-902, 6626-0811-903, 6626-0811-904, 6626-0811-905, 6626-0811-920, 6626-0811-923, 6626-0821-900, 6626-0821-903, 6626-0821-920, 6626-0911-900, 6626-0911-902, 6626-0911-903, 6626-0911-905, 6626-0911-923, 6626-0921-902, 6626-0921-903, 6626-0921-905, 6626-1011-900, 6626-1111-905, 6626-1421-905, 6626-1511-900, 6626-1521-900, 6626-1611-920, 6626-1811-900, 6628-0011-900, 6628-0011-903, 6628-0011-913, 6628-0021-903, 6628-0111-900, 6628-0111-903, 6628-0211-900, 6628-0221-900, 6628-0311-900, 6628-0321-900, 6628-0321-910, 6628-0411-900, 6628-0411-902, 6628-0411-903, 6628-0421-900, 6628-0511-900, 6628-0511-920, 6628-0511-923, 6628-0511-943, 6628-0521-943, 6628-0610-923, 6628-0611-900, 6628-0611-903, 6628-0611-905, 6628-0621-900, 6628-0711-905, 6628-0721-905, 6628-0810-923, 6628-0811-900, 6628-0811-902, 6628-0811-903, 6628-0811-905, 6628-0811-910, 6628-0811-920, 6628-0811-922, 6628-0820-923, 6628-0821-900, 6628-0821-903, 6628-0910-900, 6628-0911-900, 6628-0911-903, 6628-0911-905, 6628-0921-902, 6628-0921-920, 6628-1321-920, 6628-1421-903, 6628-1521-920, M1050214, M1051981, M1052131, M1052601, M1053655, M1053786, M1054018, M1056216, M1056217, M1061559, M1117929, M1136797, M1137588, M1137746, M1137750, QCR045; e) Care Plus 4000, Plus 4000, Model Numbers: 6600-0203-901, 6600-0225-903, 6600-0232-900, 6600-0325-901, 6600-0325-903, 6600-0326-901, 6600-0326-902, 6600-0326-903, 6600-0326-904, 6600-0326-909, 6600-0327-901, 6600-0333-901, 6600-0333-902, 6600-0334-901, 6600-0334-902, 6600-0334-903, 6600-0334-904, 6600-0334-905, 6600-0334-910, 6600-0334-911, 6600-0421-900, 6624-0111-900, 6624-0111-903, 6624-0311-900, 6624-0311-903, 6624-0321-900, 6624-0321-903, 6624-0321-943, 6624-0321-967, 6624-0411-900, 6624-0421-900, 6624-0421-905, 6624-0511-900, 6624-0511-903, 6624-0511-913, 6624-0521-900, 6624-0521-903, 6624-0611-903, 6624-0621-902, 6624-0811-900, 6624-0821-920, 6624-0821-943, 6624-0911-900, 6624-0921-900, 6624-1021-900, 6624-1311-900, 6624-1321-900, 6624-1321-903, 6624-1421-900, 6624-1421-903, 6624-1511-900, 6624-1521-900, 6627-0010-913, 6627-0011-903, 6627-0020-913, 6627-0021-900, 6627-0021-903, 6627-0021-913, 6627-0021-923, 6627-0110-923, 6627-0111-900, 6627-0111-905, 6627-0111-920, 6627-0121-900, 6627-0121-903, 6627-0121-920, 6627-0211-900, 6627-0221-900, 6627-0221-903, 6627-0221-905, 6627-0221-910, 6627-0311-900, 6627-0311-903, 6627-0311-905, 6627-0320-903, 6627-0320-913, 6627-0321-900, 6627-0321-902, 6627-0321-903, 6627-0321-910, 6627-0321-923, 6627-0321-966, 6627-0321-967, 6627-0411-900, 6627-0411-903, 6627-0421-900, 6627-0421-902, 6627-0421-903, 6627-0421-905, 6627-0421-910, 6627-0421-913, 6627-0421-920, 6627-0421-966, 6627-0421-967, 6627-0511-900, 6627-0511-903, 6627-0511-920, 6627-0511-923, 6627-0521-900, 6627-0521-903, 6627-0521-913, 6627-0521-920, 6627-0521-923, 6627-0610-923, 6627-0611-900, 6627-0611-903, 6627-0621-903, 6627-0711-900, 6627-0811-900, 6627-0811-903, 6627-0811-905, 6627-0820-913, 6627-0821-900, 6627-0821-923, 6627-0911-900, 6627-0911-903, 6627-0911-905, 6627-0911-910, 6627-0921-900, 6627-0921-903, 6627-1011-903, 6627-1021-903, 6627-1121-922, 6627-1311-900, 6627-1321-900, 6627-1321-903, 6627-1321-910, 6627-1321-966, 6627-1411-900, 6627-1411-920, 6627-1421-900, 6627-1421-903, 6627-1421-905, 6627-1421-910, 6627-1421-920, 6627-1421-953, 6627-1511-900, 6627-1921-900, 6629-0011-900, 6629-0011-905, 6629-0021-900, 6629-0021-903, 6629-0021-943, 6629-0110-923, 6629-0111-900, 6629-0121-900, 6629-0121-910, 6629-0121-923, 6629-0211-900, 6629-0221-900, 6629-0221-910, 6629-0310-923, 6629-0311-900, 6629-0311-903, 6629-0320-913, 6629-0320-923, 6629-0321-900, 6629-0321-903, 6629-0321-910, 6629-0321-913, 6629-0321-920, 6629-0321-923, 6629-0411-900, 6629-0411-903, 6629-0411-905, 6629-0420-913, 6629-0421-900, 6629-0421-902, 6629-0421-903, 6629-0421-910, 6629-0421-913, 6629-0511-900, 6629-0511-903, 6629-0521-900, 6629-0521-902, 6629-0610-923, 6629-0611-900, 6629-0611-910, 6629-0620-923, 6629-0711-900, 6629-0711-910, 6629-0810-923, 6629-0811-900, 6629-0811-902, 6629-0811-903, 6629-0811-923, 6629-0820-923, 6629-0821-900, 6629-0821-902, 6629-0821-903, 6629-0821-920, 6629-0821-923, 6629-0911-900, 6629-0920-903, 6629-0920-923, 6629-0921-900, 6629-0921-902, 6629-0921-905, 6629-1311-900, 6629-1321-900, 6629-1321-903, 6629-1411-900, 6629-1421-900, 6629-1421-903, 6629-1421-920, 6629-1511-900, M1052135, M1052600, M1053785, M1137897; ALL SERIAL NUMBERS

Quantity: 21626 units

Reason for recall

GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.

Recall record

Recall number
Z-0957-2024
Classification
Class I
Status
Ongoing
Voluntary or mandated
Voluntary: Firm initiated
Firm notification
Letter
Distribution
Worldwide distribution.
Recall initiated
2023-12-26
Classified by FDA Center
2024-02-09
FDA published
2024-02-21
Recalling firm
Wipro GE Healthcare Private Ltd.
Firm location
Bengaluru, N/A, India

Operational response

A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.

For the official FDA enforcement record, see FDA's Recall Search.