BRANMOOR
THURSDAY · 14 MAY 2026

FDA Device Recall Z-0683-2025

Nova Biomedical Corporation · Waltham, MA

Class I — life-threatening Ongoing

Highest impact — Class I device recall — FDA determination that use of, or exposure to, the device will cause serious adverse health consequences or death. Distribution is nationwide and the recall remains in progress.

Device

StatStrip Glucose Ketone (mmol/L) Hospital Meter System. Model/Catalog Number: 63910. Prescription use blood glucose ketone meter for near-patient testing

Lot / serial / GTIN: Model No 63910. UDI-DI (01)1038548063910. Software Versions: v0.0.13.10 to v0.0.13.44.

Quantity: 1,615 units

Reason for recall

A software bug within the meter firmware leads to transmission of erroneous glucose and/or ketone patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current test result to the DMS.

Recall record

Recall number
Z-0683-2025
Classification
Class I
Status
Ongoing
Voluntary or mandated
Voluntary: Firm initiated
Firm notification
E-Mail
Distribution
Distribution US Nationwide including Puerto Rico, Germany, Italy, Israel, Switzerland, Australia, Great Britain, and the Netherlands.
Recall initiated
2024-11-20
Classified by FDA Center
2024-12-20
FDA published
2025-01-01
Recalling firm
Nova Biomedical Corporation
Firm location
Waltham, MA

Operational response

A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.

For the official FDA enforcement record, see FDA's Recall Search.

‹ All device recalls