FDA Device Recall Z-0561-2024
MEDLINE INDUSTRIES, LP - Northfield · Northfield, IL
Class I — life-threatening Ongoing
Device
Tracheostomy care kits: (1) Medline Tracheostomy Clean and Care Tray with 14 Fr Suction Catheter, REF DYND40580, Sterile; (2) Medline Tracheostomy Clean and Care Tray, REF DYND4061030, Sterile; (3) Medline Basic Tracheostomy Care Kit, REF CC3T4691A, Sterile; (4) Medline Tracheostomy Care Kit with Glove, REF CC4681A, Sterile; (5) Medline Mini Tracheostomy Care Kit, REF CC3T3030A, Sterile; and (6) Medline Mini Tracheostomy Care Kit, REF DYND40622, Non-sterile.
Reason for recall
The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.
Recall record
- Recall number
Z-0561-2024- Classification
- Class I
- Status
- Ongoing
- Voluntary or mandated
- Voluntary: Firm initiated
- Firm notification
- Letter
- Distribution
- US Nationwide distribution including Puerto Rico. OUS (Foreign) distribution was made to Canada, Panama, and United Arab Emirates.
- Recall initiated
- 2023-10-16
- Classified by FDA Center
- 2023-12-22
- FDA published
- 2024-01-03
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Firm location
- Northfield, IL
Operational response
A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.
For the official FDA enforcement record, see FDA's Recall Search.