FDA Device Recall Z-0337-2022

Datascope Corp. · Mahwah, NJ

Class I — life-threatening Ongoing

Highest impact — Class I device recall — FDA determination that use of, or exposure to, the device will cause serious adverse health consequences or death. Distribution is nationwide and the recall remains in progress.

Device

Cardiosave Rescue IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-75, 0998-UC-0800-75, 0998-00-0800-83, 0998-UC-0800-83, 0998-00-0800-85, 0998-UC-0800-85.

Lot / serial / GTIN: All Serial Numbers. May also be designated as part number D998 etc. Model Number, UDI-DI: 0998-00-0800-75, 10607567112312; 0998-UC-0800-75, 10607567112312; 0998-00-0800-83, 10607567108407; 0998-UC-0800-83, 10607567108407; 0998-00-0800-85, 10607567113449; 0998-UC-0800-85, 10607567113449.

Quantity: 44 (US); 5 (OUS)

Reason for recall

Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic instability.

Recall record

Recall number
Z-0337-2022
Classification
Class I
Status
Ongoing
Voluntary or mandated
Voluntary: Firm initiated
Firm notification
Letter
Distribution
Worldwide distribution - US Nationwide.
Recall initiated
2021-11-15
Classified by FDA Center
2021-12-15
FDA published
2021-12-22
Recalling firm
Datascope Corp.
Firm location
Mahwah, NJ

Operational response

A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.

For the official FDA enforcement record, see FDA's Recall Search.