BRANMOOR
THURSDAY · 14 MAY 2026

FDA Device Recall Z-0310-2025

Cardinal Health 200, LLC · Waukegan, IL

Class I — life-threatening Ongoing

Highest impact — Class I device recall — FDA determination that use of, or exposure to, the device will cause serious adverse health consequences or death. Distribution is nationwide and the recall remains in progress.

Device

Monoject U-100 1 mL Insulin Syringe Luer-Lock with Tip Cap Soft Pack, REF 1188100777

Lot / serial / GTIN: UDI/DI 10192253034783 (EA), Lot Number 221201, exp. 2027-11-30; UDI/DI 20192253034780 (BX), Lot Number 230201, exp. 2028-01-31; UDI/DI 50192253034781 (CS), Lot Number 230202,exp. 2028-02-01

Quantity: 371,326 units

Reason for recall

Certain lots of listed lots of Cardinal Health Monoject U-100 1 mL Insulin Syringe Luer-Lock with Tip Cap Soft Pack have demonstrated incompatibility with needleless Intravenous (IV) connectors. Cardinal Health recommends that they not be used for administration of IV push insulin via needleless connector.

Recall record

Recall number
Z-0310-2025
Classification
Class I
Status
Ongoing
Voluntary or mandated
Voluntary: Firm initiated
Firm notification
Letter
Distribution
Nationwide
Recall initiated
2024-09-25
Classified by FDA Center
2024-11-08
FDA published
2024-11-20
Recalling firm
Cardinal Health 200, LLC
Firm location
Waukegan, IL

Operational response

A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.

For the official FDA enforcement record, see FDA's Recall Search.