BRANMOOR
THURSDAY · 14 MAY 2026

FDA Device Recall Z-0208-2025

Percussionaire Corporation · Sandpoint, ID

Class I — life-threatening Ongoing

Highest impact — Class I device recall — FDA determination that use of, or exposure to, the device will cause serious adverse health consequences or death.

Device

Phasitron5 Breathing Circuit (P5-10, P5-HC, P5-UC-10). component of Intrapulmonary Percussive Ventilation, used to provide airway clearance therapy

Lot / serial / GTIN: Model Numbers: P5-10 P5-HC P5-UC-10 UDI-DI codes: Lot Numbers: 240321 240329 240412 240430 240509 240516 240614 240625 240627

Quantity: 43,830 devices

Reason for recall

As a result of product complaints, breathing circuits are experiencing discolored fluid which are formed when therapy was being given with specific medication through the nebulizer cup. Discoloration is caused by a reaction between medicated mist and an inadvertent/unapproved addition of a nickel coating on the spring component.

Recall record

Recall number
Z-0208-2025
Classification
Class I
Status
Ongoing
Voluntary or mandated
Voluntary: Firm initiated
Firm notification
Letter
Distribution
U.S.: PA, AR, MI, CA, TX, OH, NY, FL, IL, NE, OR, GA, WI, MO, OK, AL, WV, MT, CO, LA, VA, CT, SC, NH, NV, NC, NJ, MI, UT, ID, ME, MN, SD, ID, ND, KS, KY, TN, MD, VT, MA, RI, O.U.S.: Canada, Switzerland (all products shipped OUS remained in Sentec AG's (distributor, parent company of Percussionaire) control, were quarantined, and did not leave the warehouse).
Recall initiated
2024-09-16
Classified by FDA Center
2024-10-30
FDA published
2024-11-06
Recalling firm
Percussionaire Corporation
Firm location
Sandpoint, ID

Operational response

A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.

For the official FDA enforcement record, see FDA's Recall Search.

‹ All device recalls