BRANMOOR
THURSDAY · 14 MAY 2026

FDA Device Recall Z-0191-2024

Olympus Corporation of the Americas · Center Valley, PA

Class I — life-threatening Ongoing

Highest impact — Class I device recall — FDA determination that use of, or exposure to, the device will cause serious adverse health consequences or death. Distribution is nationwide and the recall remains in progress.

Device

EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T180, BF-1TQ180, BF-P180

Lot / serial / GTIN: Model BF-1T180 UDI-DI: N/A; Model BF-1TQ180 UDI-DI: N/A; Model BF-P180 UDI-DI: N/A;

Quantity: 13950 devices

Reason for recall

There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.

Recall record

Recall number
Z-0191-2024
Classification
Class I
Status
Ongoing
Voluntary or mandated
Voluntary: Firm initiated
Firm notification
Letter
Distribution
US Nationwide distribution
Recall initiated
2023-10-12
Classified by FDA Center
2023-11-09
FDA published
2023-11-15
Recalling firm
Olympus Corporation of the Americas
Firm location
Center Valley, PA

Operational response

A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.

For the official FDA enforcement record, see FDA's Recall Search.