FDA Device Recall Z-0161-2026

IMPRIS Imaging Inc · Chaska, MN

Class I — life-threatening Ongoing

Highest impact — Class I device recall — FDA determination that use of, or exposure to, the device will cause serious adverse health consequences or death. Distribution is nationwide and the recall remains in progress.

Device

IMRIS Operating Suite, Neuro III-SV iMRI System, iMRX System, containing Siemens 3 Tesla MRI Magnet, Verio

Lot / serial / GTIN: IMRIS Part Number(s): 105118-000, UDI/DI None predates UDI; 105118-001, UDI/DI None predates UDI; 105117-003, UDI/DI None predates UDI; 105117-004, UDI/DI None predates UDI; 105048-000, UDI/DI None predates UDI; 105048-001, UDI/DI None predates UDI; 105103-001, UDI/DI None predates UDI; 105071-004, UDI/DI None predates UDI; 105824-000, UDI/DI 00857534006189; 105828-000, UDI/DI 00857534006165; 116189-000, UDI/DI 00857534006202; 118396-000, UDI/DI 00857534006196.

Quantity: N/A

Reason for recall

There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

Recall record

Recall number
Z-0161-2026
Classification
Class I
Status
Ongoing
Voluntary or mandated
Voluntary: Firm initiated
Firm notification
E-Mail
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Sweden, Switzerland, Japan, Australia, Qatar, and China.
Recall initiated
2025-09-30
Classified by FDA Center
2025-10-28
FDA published
2025-11-05
Recalling firm
IMPRIS Imaging Inc
Firm location
Chaska, MN

Operational response

A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.