FDA Device Recall Z-0046-2025

Baxter Healthcare Corporation · Deerfield, IL

Class I — life-threatening Ongoing

Highest impact — Class I device recall — FDA determination that use of, or exposure to, the device will cause serious adverse health consequences or death. Distribution is nationwide and the recall remains in progress.

Device

Welch Allyn Life2000 Ventilation System, REF MS-01-0118, within exterior carton codes: 1) BT-20-0002, 2) BT-20-0002A, 3) BT-20-0002AP, 4) BT-20-0007, 5) BT-20-0011, and 6) RMS010118CP

Lot / serial / GTIN: UDI/DI 00887761978089 or 00815410020537, All Life2000 Ventilators with software version 06.08.00.00

Quantity: 10,957 units

Reason for recall

The Life2000 ventilator may fail to initiate the Low Gas Pressure alarm if the pressure gas source (Life2000 compressor, oxygen cylinder or wall source) is not supplied to the ventilator before initiating therapy.

Recall record

Recall number
Z-0046-2025
Classification
Class I
Status
Ongoing
Voluntary or mandated
Voluntary: Firm initiated
Firm notification
Letter
Distribution
US nationwide distribution.
Recall initiated
2024-09-12
Classified by FDA Center
2024-10-22
FDA published
2024-10-30
Recalling firm
Baxter Healthcare Corporation
Firm location
Deerfield, IL

Operational response

A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.

For the official FDA enforcement record, see FDA's Recall Search.