FDA Device Recall Z-0028-2026

NOXBOX LTD

Class I — life-threatening Ongoing

Highest impact — Class I device recall — FDA determination that use of, or exposure to, the device will cause serious adverse health consequences or death.

Device

NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I

Lot / serial / GTIN: Model Numbers: (1) NOXBOXI (NBL (Manufacturer)), (2) REQNOXBOXI (Linde (U.S. Distributor)); UDI-DI: 05060541640009; All serial numbers.

Quantity: 1667 units

Reason for recall

Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).

Recall record

Recall number
Z-0028-2026
Classification
Class I
Status
Ongoing
Voluntary or mandated
Voluntary: Firm initiated
Firm notification
E-Mail
Distribution
Domestic: TN
Recall initiated
2025-09-09
Classified by FDA Center
2025-10-06
FDA published
2025-10-15
Recalling firm
NOXBOX LTD

Operational response

A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.

For the official FDA enforcement record, see FDA's Recall Search.