FDA Device Recall Z-0028-2026
NOXBOX LTD
Class I — life-threatening Ongoing
Device
NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I
Reason for recall
Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).
Recall record
- Recall number
Z-0028-2026- Classification
- Class I
- Status
- Ongoing
- Voluntary or mandated
- Voluntary: Firm initiated
- Firm notification
- Distribution
- Domestic: TN
- Recall initiated
- 2025-09-09
- Classified by FDA Center
- 2025-10-06
- FDA published
- 2025-10-15
- Recalling firm
- NOXBOX LTD
Operational response
A Class I device recall indicates a strong likelihood of serious adverse health consequence or death from continued use. Identify affected units by serial number, lot, or GTIN against your inventory and against implanted-device patient registry. Pull affected inventory from active use immediately. For implanted devices, follow the recalling firm’s patient-notification protocol; in most cases this requires informing affected patients and their treating physicians directly.
For the official FDA enforcement record, see FDA's Recall Search.